ARTiCAT

CBCT tomograph for the hand district

ARTiCAT is a high resolution Computerized Tomograph (CBTC) intended for the volumetric reconstruction of the hand and wrist. It belongs to the category of so-called HRpCT (High Resolution peripheral
Computed Tomography) or ‘in vivo’ microCT.

ARTiCAT is characterized by high spatial resolution, speed of execution and ease of use.

The high resolution value makes this HRpCT particularly suitable for the visualization of micro-erosions, microneoformations,
alterations of the trabecular structure.

ARTiCAT allows you to choose between four different examination protocols for FOV (Field Of View), spatial resolution and scan duration.

Risoluzione spaziale VS FOV
ProtocolNameFOV(1) (𝜱 x h) (mm)mAsvoxel(μm)MTF 10%(μm)Tscan(s)CTDI Vol
1Large Field97x974,8135171228,7 mGy
2Large Field HR64x644,865112368,7 mGy
3Zoom83x834,8100149226,6 mGy
4Zoom HR51x514,85082366,6 mGy
Protocol1234
NameLarge FieldLarge Field HR Zoom Zoom HR
FOV(1) (𝜱 x h) (mm) 97x97 64x6483x8351x51
mAs4,84,84,84,8
voxel(μm) 13565 10050
MTF 10%(μm) 17111214982
Tscan(s) 22362236
CTDI Vol
8,7 mGy8,7 mGy6,6 mGy6,6 mGy